• Anaesthesia · May 1995

    Randomized Controlled Trial Clinical Trial Controlled Clinical Trial

    Evaluation of the influence of movement on saturation readings from pulse oximeters.

    • J L Plummer, A Z Zakaria, A H Ilsley, R R Fronsko, and H Owen.
    • Department of Anaesthesia and Intensive Care, Flinders Medical Centre, Bedford Park, South Australia.
    • Anaesthesia. 1995 May 1;50(5):423-6.

    AbstractThis study aimed to develop a protocol for assessing the influence of movement on oxyhaemoglobin saturation readings from pulse oximeters. Thirty-six volunteers took part in the study. In each volunteer, each hand was monitored by both a Nellcor N200 oximeter using a disposable probe and by a Datex Satlite DS103 oximeter using a clip-on finger probe. Volunteers made five standardised movements during which output was recorded from all four oximeters. All movements were associated with apparent decreases in oxyhaemoglobin saturation which were statistically significant for two movements with the Nellcor equipment and for four movements with the Datex equipment. Movement was associated with increases in the magnitude of pulse amplitude, but this was not quantitatively associated with magnitude of artefactual changes in saturation. Use of this standardised movement protocol allows quantification of movement artefact from pulse oximeters and should facilitate the development of equipment less affected by movement.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…