Journal of nuclear medicine : official publication, Society of Nuclear Medicine
-
Multicenter Study
Qualification of PET scanners for use in multicenter cancer clinical trials: the American College of Radiology Imaging Network experience.
The PET Core Laboratory of the American College of Radiology Imaging Network (ACRIN) qualifies sites to participate in multicenter research trials by quantitatively reviewing submitted PET scans of uniform cylinders to verify the accuracy of scanner standardized uptake value (SUV) calibration and qualitatively reviewing clinical PET images from each site. To date, cylinder and patient data from 169 PET scanners have been reviewed, and 146 have been qualified. ⋯ Minimizing errors in SUV measurement is critical to achieving accurate quantification in clinical trials. The experience of the ACRIN PET Core Laboratory shows that many sites are unable to maintain accurate SUV calibrations without additional training or supervision. This raises concerns about using SUVs to quantify patient data without verification.