Coagulopathic bleeding during and after cardiac surgery is associated with increased morbidity and mortality. Viscoelastic testing is increasingly used instead of laboratory testing. Our aim was to compare a new viscoelastic point-of-care device, the Quantra® System, with thromboelastography and standard laboratory testing. ⋯ These Quantra thresholds showed high negative predictive value for low platelets (platelet clot stiffness, 97.4%), prolonged activated partial thromboplastin time (clot time, 92.6%) and reduced maximum amplitude on thromboelastography (clot stiffness, 93.6%). The Quantra predicted clinical need for transfusion of platelets (area under the curve 0.71, p = 0.001) but all tests performed poorly at predicting the need for fresh frozen plasma transfusion. We have shown that point-of-care testing using the novel Quantra system provides useful data for guiding transfusion management.
-
Obstructive sleep apnoea and residual neuromuscular blockade are, independently, known to be risk factors for respiratory complications after major surgery. Residual effects of neuromuscular blocking agents are known to reduce the hypoxic ventilatory response in healthy volunteers. ⋯ The hypoxic ventilatory response was reduced by 32% (p = 0.016) during residual neuromuscular block (rocuronium to train-of-four ratio 0.7), but the hypercapnic ventilatory response was unaffected. We conclude that neuromuscular blockade specifically depresses peripheral chemosensitivity, and not respiratory muscle function since the hypercapnic ventilatory response was unaffected.