Analytical biochemistry
-
Analytical biochemistry · Jun 1990
Comparative StudyAutomated enzymatic measurement of adenosine deaminase isoenzyme activities in serum.
We developed a simple, rapid, and automated method for simultaneous measurement of adenosine deaminase (ADA, EC 3.5.4.4) isoenzymes in human serum, based on their apparent difference in Ki values for erythro-9-(2-hydroxy-3-nonyl)adenine (EHNA) as inhibitor. Serum ADA was partially purified by CM-Sephadex, gel-filtration, and affinity chromatography into two types of isoenzymes, designated ADA1 (300 kDa) and ADA2 (120 kDa). Because ADA2 has a higher Km for adenosine and higher Ki values for EHNA than does ADA1, the activity of ADA1 is almost completely inhibited by EHNA at 0.1 mM (analytical recovery 4.1%), whereas ADA2 is practically unaffected (analytical recovery 94.8%) by that concentration of EHNA. ⋯ Results of the present method correlated well (r = 0.9026 for ADA1, 0.9438 for ADA2) with those of the ion-exchange chromatography method. The upper limits of the reference intervals, as calculated from data for 320 healthy donors, are 7.2 U/liter for ADA1, and 14.6 U/liter for ADA2. This method is suitable for analysis of large numbers of samples in clinical laboratories for routine monitoring of the activities of ADA isoenzymes in serum.