The effects of ventricular drainage on the intracranial pressure signal and the pressure reactivity index.
In subarachnoid hemorrhage (SAH) patients intracranial pressure (ICP) is usually monitored via an extraventricular drain (EVD), which can produce false readings when the drain is open. It is established that both the ICP cardiac pulse frequency and long term trends over several hours are often seriously corrupted. The aim of this study was to establish whether or not the intermediate frequency bands [respiratory, Mayer wave and very low frequency (VLF)] were also corrupted. ⋯ We found a small but statistically significant decrease during open-drain periods. Based on analysis of the change in the PRx distribution during open drainage we concluded that this decrease also represented physiological changes rather than artifact. In summary the ICP respiratory, Mayer wave, and VLF frequency bands are not corrupted when the EVD is open, and it safe to use these for autoregulation studies.
-
J Clin Monit Comput · Apr 2017
Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position.
To guarantee accurate measurement of central venous pressure (CVP) or pulmonary artery occlusion pressure (PAOP), proper positioning of a reference transducer is a prerequisite. We investigated ideal transducer levels in supine, prone, and sitting position in adults. Chest computed tomography images of 113 patients, taken in supine or prone position were reviewed. ⋯ The ratio of the most cephalad blood level of RA and LA to the sternal length was 0.70 ± 0.10 and 0.68 ± 0.09 from the mid-sternoclavicular joint in sitting position, which corresponded to the upper border of 4th rib. Optimal CVP transducer levels are at four-fifths of the AP diameter of thorax in supine position, at a half of that in prone position, and at upper border of the 4th sternochondral joint in sitting position. PAOP transducer levels are similar in prone and sitting position, except for supine position which is at three-fifths of the AP diameter of thorax.