• Arch Orthop Trauma Surg · Jun 2024

    The peak force to push a trial tibial insert into position cannot be used to select the correct thickness in total knee arthroplasty.

    • Gabriel Santana, S M Howell, and M L Hull.
    • Department of Biomedical Engineering, University of California Davis, Davis, CA, 95616, US.
    • Arch Orthop Trauma Surg. 2024 Jun 1; 144 (6): 276727732767-2773.

    AbstractAn objective of a total knee arthroplasty (TKA) is to restore native (i.e. healthy) function, and a crucial step is determining the correct insert thickness for each patient. If the insert is too thick, then stiffness results, and if too thin, then instability results. Two methods to determine the insert thickness are by manually assessing the joint laxity and by using a trial insert with goniometric markings that measures the internal-external rotation of the trial with respect to a mark on the femoral component. The former is qualitative and depends on the surgeon's experience and 'feel' and while the latter is quantitative, it can be used only with an insert with medial ball-in-socket conformity. An unexplored method is to measure the force required to push a trial insert into position. To determine whether this method has merit, the push force was measured in 30 patients undergoing unrestricted kinematically aligned TKA using an insert with ball-in-socket medial conformity, a flat lateral surface, and retention of the posterior cruciate ligament. During surgery, the surgeon determined three appropriate thicknesses to test from a selection ranging from 10 mm to 14 mm in 1 mm increments. The peak push forces going from an insert 1 mm thinner than the correct thickness as determined by an insert goniometer and from the correct thickness to 1 mm thicker were measured. Mean peak forces for the different insert thicknesses were 127 ± 104 N, 127 ± 95 N, and 144 ± 96 N for 1 mm thinner, correct, and 1 mm thicker, respectively, and did not differ (p = 0.3210). As a result, measurement of peak force during trial positioning of a tibial insert cannot be used to identify the correct thickness for all insert designs.© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.