IEEE transactions on bio-medical engineering
-
The strength-duration curve is a plot of the threshold current (I) versus pulse duration (d) required to stimulate excitable tissue. On this curve are two points: 1) rheobase (b) and 2) chronaxie (c). Rheobase is the threshold current for an infinitely long-duration stimulus. ⋯ The mathematical expression for the strength-duration curve is I = b(1 + c/d). Although there are many published values for chronaxie for various excitable tissues, the range of variability for a given tissue type is quite large. This paper identifies five factors that can affect the accuracy of chronaxie measurement and shows that the most reliable values can be obtained with a rectangular pulse delivered from a constant-current source.
-
IEEE Trans Biomed Eng · Sep 1992
ReviewRegeneration microelectrode array for peripheral nerve recording and stimulation.
A microelectrode array capable of recording from and stimulating peripheral nerves at prolonged intervals after surgical implantation has been demonstrated. The microelectrode array, fabricated on a silicon substrate perforated by multiple holes (referred to as via holes), is implanted between the ends of a surgically severed nerve. ⋯ All fabrication methods were designed to be compatible with standard CMOS/BiCMOS processes to allow for on-chip signal processing circuits in future designs. Such arrays, implanted in the peroneal nerves of rats, were used to record from and stimulate the nerves at up to 13 months postoperatively.