Circulation
-
Comparative Study Clinical Trial Controlled Clinical Trial
Defibrillation efficacy of commercially available biphasic impulses in humans. Importance of negative-phase peak voltage.
Recent studies have shown that specifically shaped biphasic waveforms can lower energy requirements for ventricular defibrillation. We prospectively compared the defibrillation efficacy of three different biphasic wave shapes incorporated in three commercially available implantable defibrillators. The results led to the development of a second protocol in which the importance of negative-phase peak voltage and duration was investigated. ⋯ Our results suggest that specifically shaped biphasic waveforms delivered from commercially available devices can affect energy requirements for defibrillation. More importantly, the amplitude of the P2 peak voltage may be a more critical determinant than the P2 duration for defibrillation success of biphasic waveforms in humans.