Journal of nuclear medicine : official publication, Society of Nuclear Medicine
-
The goal of this study was to describe our initial experience with the deep-inspiration breath-hold (DIBH) technique in combined PET/CT of the thorax. This article presents particular emphasis on the technical aspects required for clinical implementation. ⋯ The practice of DIBH PET/CT is feasible in a clinical setting. With this technique, consistent lung inflation levels are achieved during PET/CT sessions, as judged by both motion tracker and verification of spatial matching between PET and CT images. Breathing-induced motion artifacts are significantly reduced using DIBH compared with free breathing, enabling better target localization and quantitation. The DIBH technique showed an increase in the median SUV by 32.46%, with a range from 4% to 83%, compared with SUVs measured on the clinical images. The median percentage reduction in the PET-to-CT lesions' centroids was 26.6% (range, 3%-50%).