Evaluation of membrane oxygenators and reservoirs in terms of capturing gaseous microemboli and pressure drops.
An increasing amount of evidence points to cerebral embolization during cardiopulmonary bypass (CPB) as the principal etiologic factor of neurologic complications. In this study, the capability of capturing and classification of gaseous emboli and pressure drop of three different membrane oxygenators (Sorin Apex, Terumo Capiox SX25, Maquet QUADROX) were measured in a simulated adult model of CPB using a novel ultrasound detection and classification quantifier system. The circuit was primed with 1000 mL heparinized human packed red blood cells and 1000 mL lactated Ringer's solution (total volume 2000 mL, corrected hematocrit 26-28%). ⋯ Microemboli counts uniformly increased with hypothermic perfusion (25 degrees C). Different types of oxygenators and reservoirs have different capability of capturing gaseous emboli and transmembrane pressure drop. Based on this investigation, Maquet QUADROX membrane oxygenator has the lowest pressure drop and better capability for capturing gaseous microemboli.
-
Due to improved outcome after surgery for congenital heart defects, children, adolescents, and grown-ups with congenital heart defects become an increasing population. In order to evaluate operative risk and early outcome after mechanical aortic valve replacement (AVR) in this population, we reviewed patients who underwent previous repair of congenital heart defects. Between July 2002 and November 2008, 15 (10 male and 5 female) consecutive patients (mean age 14.5 +/- 10.5 years) underwent mechanical AVR. ⋯ At the latest clinical evaluation, all patients were in good clinical condition without a pathological increased gradient across the aortic valve prosthesis or paravalvular leakage in echocardiography. Mechanical AVR has excellent results in patients after previous repair of congenital heart defects in childhood, even in combination with complex concomitant procedures. Previous operations do not significantly affect postoperative outcome.
-
The nonpulsatile blood flow obtained using standard cardiopulmonary bypass (CPB) circuits is still generally considered an acceptable, nonphysiologic compromise with few disadvantages. However, numerous reports have concluded that pulsatile perfusion during CPB achieves better multiorgan response postoperatively. ⋯ Therefore, we considered that optimizations of CPB unit and extracorporeal life support (ECLS) system circuit components were needed to deliver sufficient pulsatile flow. In addition, energy equivalent pressure, surplus hemodynamic energy, and total hemodynamic energy, calculated using pressure and flow waveforms, were used to evaluate the pulsatilities of pulsatile CPB and ECLS systems.
-
Comparative Study
Comparison of pumps and oxygenators with pulsatile and nonpulsatile modes in an infant cardiopulmonary bypass model.
As the evidence mounts in favor of pulsatile perfusion during CPB, it is necessary to investigate the effect of circuit components on the quality of pulsatility delivered throughout the circuit. We compared two bloodpumps, the Jostra HL-20 heart-lung machine and the MEDOS DELTASTREAM DP1 Bloodpump, and two oxygenators, the Capiox Baby RX05 and the MEDOS HILITE 800LT, in terms of mean arterial pressure, energy equivalent pressure, surplus hemodynamic energy, total hemodynamic energy, and pressure drop over the oxygenators using a blood analog. The pumps and oxygenators were combined in unique circuits and tested in nonpulsatile and pulsatile modes, at two flow rates (500 and 800 mL/min), and three rotational speed differentials when using the MEDOS DELTASTREAM DP1 Bloodpump for 144 trials in total. ⋯ The MEDOS DP1 Bloodpump produced almost no pulsatility in nonpulsatile mode. Pressure drops over the Capiox Baby RX05 were markedly higher, at 92.5 +/- 0.4 mm Hg with the MEDOS DP1 Bloodpump at 800 mL/min and 4500 rpm in pulsatile mode, than those of the MEDOS HILITE 800LT oxygenator, which was 67.0 +/- 0.1 mm Hg at the same settings. These results suggest that careful selection of each circuit component, based on the individual clinical case and component specifics, are necessary to achieve the best quality of pulsatility.
-
This study investigated features and treatments of perioperative coagulopathies in cyanotic infants with complex congenital heart disease (CCHD). Thirty-six infants with cyanotic CCHD were involved and divided into two groups: In group H (n = 20), hematocrit (HCT) > 54%, and in group L (n = 16), HCT < 54%. Blood was sampled at anesthesia induction (T1), rewarming to 36 degrees C (T2), after heparin neutralization (T3), and 4 h after operation (T4). ⋯ After therapy, PLT function in both groups restored to T1 level (P > 0.05); Ffg at T4 was significantly better than at T1 (P < 0.01) in group H, but Ffg at T4 with still normal function was lower than at T1 in group L (P < 0.01). Whole hemostatic function at T4 was back to normal and had no differences between two groups. So, we proposed that fibrinogen and PLT transfusion in combination should be better for infants with high HCT CCHD, but PLT alone might be enough for low HCT ones.