Resuscitation
-
The purpose of this study was to measure the local electrical field or potential gradient, measured with a catheter-based system, required to terminate long duration electrically or ischaemically induced ventricular fibrillation (VF). We hypothesized that prolonged ischaemic VF would be more difficult to terminate when compared to electrically induced VF of similar duration. ⋯ Defibrillation of prolonged VF produced by acute myocardial ischaemia requires a significantly greater potential gradient to terminate than prolonged VF induced by electrical stimulation of the right ventricular endocardium. The VF duration used in this study approximates that occurring in victims of out-of-hospital cardiac arrest. Our findings may be of clinical importance in the management of such patients.