Continuous central venous oxygen saturation monitoring under varying physiological conditions in an animal model.
We compared saturations from a paediatric central venous oximetry catheter with co-oximetry values with changes in drug infusions, intravascular blood volume and hypoxia in an animal model. Piglets (large white) were anaesthetised, intubated and mechanically ventilated. PediaSat oximetry catheters were placed in the superior vena cava via jugular vein cut-down and in the inferior vena cava percutaneously via the femoral vein. ⋯ During the hypoxia study there was good agreement between the co-oximeter to ScvO2-superior catheter but poor agreement with to the inferior vena cava catheter samples. In the hypovolaemic phase of the experiment there was good agreement between the measured co-oximetry value and ScvO2-superior catheter until the mean blood pressure reached 43 mmHg. The oximetry catheter is capable of identifying changes in ScvO2 under physiological conditions usually encountered in clinical medicine but was less accurate at the extremes of physiology and when placed in the inferior vena cava catheter especially during hypovolaemia and hypoxia.