Evaluation of the new generation dual-lumen catheter for neonatal ECMO.
The purpose of this study was to compare the newly designed dual-lumen venovenous catheter (VR13, OriGen Biomedical, Austin, TX) with the current dual-lumen catheter (VV12, OriGen Biomedical). ⋯ These findings indicate that the VR13 catheter resulted in levels of recirculation equal to the VV12. Based on resistance measurements, we do not recommend the use of this new catheter beyond 400 ml/min until minor design changes are made.
-
Jehovah's Witnesses present a challenge to cardiac surgeons, as quality of care is not only defined by mortality and morbidity, but also by the avoidance of blood transfusions. Over the last years, minimized perfusion circuits (MPC) have contributed substantially to the achievement of this goal in our clinic. Presented is a retrospective analysis of our experience. ⋯ These encouraging results emphasize the role MPCs can play in optimizing the quality of patient care. We hope that this report can serve as a stimulus for similar experiences.