Breathing filters or heat and moisture exchangers (HME), which are placed between a tracheal tube and the Y-piece, allow reuse of breathing tubes without changing between operations. During low-flow and minimal-flow anaesthesia, condensed water accumulates in the breathing circuit. An evaluation of the volume of condensed water is given (Fig. 1). ⋯ For the experiments, a commercial breathing circuit and ventilator (Dräger Sulla 808 + Ventilog) and patient model (Fig. 2) were used; 12 breathing filters/HMEs of different manufacturers were tested. Only 3 filters were not permeable to the test volume of 20 ml water. The authors suggest the water volume be checked routinely and the breathing tubes be emptied if necessary.