British journal of anaesthesia
-
Randomized Controlled Trial
Influence of variations in systemic blood flow and pressure on cerebral and systemic oxygen saturation in cardiopulmonary bypass patients.
Although both pressure and flow are considered important determinants of regional organ perfusion, the relative importance of each is less established. The aim of the present study was to evaluate the impact of variations in flow, pressure, or both on cerebral and whole-body oxygen saturation. ⋯ In the present elective cardiac surgery population, Sc(O₂) and Sv(O₂) were significantly lower with lower flow, regardless of systemic arterial pressure. Moreover, phenylephrine administration was associated with a reduced cerebral and systemic oxygen saturation.
-
Multicenter Study Clinical Trial
Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients.
A new calibrated pulse wave analysis method (VolumeView™/EV1000™, Edwards Lifesciences, Irvine, CA, USA) has been developed to continuously monitor cardiac output (CO). The aim of this study was to compare the performance of the VolumeView method, and of the PiCCO2™ pulse contour method (Pulsion Medical Systems, Munich, Germany), with reference transpulmonary thermodilution (TPTD) CO measurements. ⋯ The performance of the new VolumeView™-CCO method is as reliable as the PiCCO2™-CCO pulse wave analysis in critically ill patients. However, an improved precision was observed with the VolumeView™ technique. CLINICALTRIALS.GOV IDENTIFIER: NCT01405040.
-
Editorial Review
Perioperative database research: possibilities and pitfalls.