Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery.
Accuracy and performance of the only currently available intra-arterial blood-gas monitoring system (Paratrend 7, PT7) were assessed in 23 patients during thoracoscopic surgery using one-lung ventilation. Over a wide range of values for arterial PO2 (6.1-61.1 kPa), PCO2 (4.1-9.5 kPa) and pH (7.19-7.50), 138 arterial blood-gas values obtained by PT7 were compared with corresponding in vitro laboratory blood-gas measurements. ⋯ Also, the bias for sequential changes between two, consecutive times was not significantly different from the ideal value of 0. We conclude that the PT7 is helpful in monitoring patients during thoracoscopy.
-
In an open study we have investigated the efficacy of 20 ml of 0.75% ropivacaine (7.5 mg ml-1) to provide extradural anaesthesia for elective Caesarean section. Plasma concentrations (total and free) were estimated in the mother and neonate. ⋯ Plasma concentrations of ropivacaine in the mother and neonate were within safe limits and consistent with previous studies. Two mothers received accidental i.v. injections of ropivacaine 75 mg and 150 mg, respectively, without serious adverse effects.