The Medical journal of Australia
-
The accuracy and reliability of new blood glucose measurement strips, BM Test-Glycemie 20-800, were assessed in 99 consecutive patients attending the diabetic clinic and seven subjects at home. The BM Test-Glycemie 20-800 strip technique was compared with the automated hexokinase method and the Reflotest-Glucose test strip technique employing the Boehringer Reflomat. ⋯ Importantly, the strips were often misleading in indicating hypoglycaemia. It is concluded that the BM Test-Glycemie 20-800 strips have a practical and valuable role when used by suitably instructed staff members, or by patients in the management of diabetes, and represent a major advance over urine testing on previously available direct reading enzyme strip methods.