Respiratory care
-
The performance of high-frequency oscillatory ventilators (HFOV) differs by the waveform generation mode and circuit characteristics. Few studies have described the performance of piston-type HFOV. The present study aimed to compare the amplitude required to reach the target high-frequency tidal volume ([Formula: see text]); determine the relationship between the settings and actual pressure in amplitude or mean airway pressure ([Formula: see text]); and describe the interaction among compliance, frequency, and endotracheal tube (ETT) inner diameter in 4 HFOV models, including Humming X, Vue (a piston type ventilator commonly used in Japan), VN500 (a diaphragm type), and SLE5000 (a reverse jet type). ⋯ The actual measured value, such as alveolar [Formula: see text] and high-frequency [Formula: see text], varied according to the type of HFOV system and the inner diameter of the ETT, even with identical settings. Clinicians should therefore determine the setting appropriate to each HFOV model.