• N. Engl. J. Med. · Dec 2012

    Randomized Controlled Trial Multicenter Study Comparative Study

    Reduction in inappropriate therapy and mortality through ICD programming.

    • Arthur J Moss, Claudio Schuger, Christopher A Beck, Mary W Brown, David S Cannom, James P Daubert, N A Mark Estes, Henry Greenberg, W Jackson Hall, David T Huang, Josef Kautzner, Helmut Klein, Scott McNitt, Brian Olshansky, Morio Shoda, David Wilber, Wojciech Zareba, and MADIT-RIT Trial Investigators.
    • Department of Medicine, University of Rochester Medical Center, Rochester, NY 14642-0653, USA. heartajm@heart.rochester.edu
    • N. Engl. J. Med.. 2012 Dec 13;367(24):2275-83.

    BackgroundThe implantable cardioverter-defibrillator (ICD) is highly effective in reducing mortality among patients at risk for fatal arrhythmias, but inappropriate ICD activations are frequent, with potential adverse effects.MethodsWe randomly assigned 1500 patients with a primary-prevention indication to receive an ICD with one of three programming configurations. The primary objective was to determine whether programmed high-rate therapy (with a 2.5-second delay before the initiation of therapy at a heart rate of ≥200 beats per minute) or delayed therapy (with a 60-second delay at 170 to 199 beats per minute, a 12-second delay at 200 to 249 beats per minute, and a 2.5-second delay at ≥250 beats per minute) was associated with a decrease in the number of patients with a first occurrence of inappropriate antitachycardia pacing or shocks, as compared with conventional programming (with a 2.5-second delay at 170 to 199 beats per minute and a 1.0-second delay at ≥200 beats per minute).ResultsDuring an average follow-up of 1.4 years, high-rate therapy and delayed ICD therapy, as compared with conventional device programming, were associated with reductions in a first occurrence of inappropriate therapy (hazard ratio with high-rate therapy vs. conventional therapy, 0.21; 95% confidence interval [CI], 0.13 to 0.34; P<0.001; hazard ratio with delayed therapy vs. conventional therapy, 0.24; 95% CI, 0.15 to 0.40; P<0.001) and reductions in all-cause mortality (hazard ratio with high-rate therapy vs. conventional therapy, 0.45; 95% CI, 0.24 to 0.85; P=0.01; hazard ratio with delayed therapy vs. conventional therapy, 0.56; 95% CI, 0.30 to 1.02; P=0.06). There were no significant differences in procedure-related adverse events among the three treatment groups.ConclusionsProgramming of ICD therapies for tachyarrhythmias of 200 beats per minute or higher or with a prolonged delay in therapy at 170 beats per minute or higher, as compared with conventional programming, was associated with reductions in inappropriate therapy and all-cause mortality during long-term follow-up. (Funded by Boston Scientific; MADIT-RIT ClinicalTrials.gov number, NCT00947310.).

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.