• J Clin Monit Comput · Feb 2014

    Comparative Study

    Detection of changes in muscle oxygen saturation in the human leg: a comparison of two near-infrared spectroscopy devices.

    • Andreas Nygren, Kajsa Rennerfelt, and Qiuxia Zhang.
    • Department of Cardiothoracic Anaesthesia and Intensive Care, Sahlgrenska University Hospital, Göteborg, Sweden, andreas.nygren@aniv.gu.se.
    • J Clin Monit Comput. 2014 Feb 1;28(1):57-62.

    AbstractThe purpose of this study was first to evaluate the near infrared spectroscopy (NIRS) device, INVOS 4100 as a method of measuring exercise and arterial occlusion induced muscle ischemia in human leg, by comparison with InSpectra tissue spectrometer Model 325, and secondly to determine the influence of skin and subcutaneous thickness on the NIRS measurements. Twenty healthy subjects (43 ± 8 years) volunteered for the study. Tissue oxygen saturation (StO2) in the anterior tibial muscle was measured simultaneously with InSpectra Model 325 in one leg and INVOS 4100 in the contralateral leg during an exercise test until muscle exhaustion and arterial occlusion with and without exercise. The skin and subcutaneous thickness was identified by ultrasound imaging. Baseline StO2 was 87 ± 8 % detected by InSpectra and 76 ± 6 % by INVOS. Both devices detected an immediate decrease of StO2 (p < 0.001) during exercise, arterial occlusion with and without exercise, and a significant post ischemic hyperaemia (p < 0.001) during recovery. There was a significant inverse correlation between skin and subcutaneous thickness and baseline StO2 (r = -0.78, p < 0.01) as well as change in StO2 during exercise (r = -0.65, p = 0.002) for InSpectra, which was not apparent for INVOS. The results demonstrate that the cerebral/somatic oxygenation monitor (INVOS) is able to detect experimentally induced skeletal muscle ischemia in the human leg as well as the peripheral tissue spectrometer (InSpectra). Muscle oxygen saturation measurement by INVOS is less affected by skin and subcutaneous thickness than measured by InSpectra.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…