• Artificial organs · Jan 2019

    Effects of Pulsatile Control Algorithms for Diagonal Pump on Hemodynamic Performance and Hemolysis.

    • Shigang Wang, Madison Force, Morgan K Moroi, Sunil Patel, Allen R Kunselman, and Akif Ündar.
    • Penn State Hershey Pediatric Cardiovascular Research Center, Department of Pediatrics, Penn State Milton S. Hershey Medical Center, Penn State Hershey College of Medicine, Penn State Hershey Children's Hospital, Hershey, PA, USA.
    • Artif Organs. 2019 Jan 1; 43 (1): 60-75.

    AbstractThe objective of this study is to compare hemodynamic performances under different pulsatile control algorithms between Medos DeltaStream DP3 and i-cor diagonal pumps in simulated pediatric and adult ECLS systems. An additional pilot study was designed to test hemolysis using two pumps during 12h-ECLS. The experimental circuit consisted of parallel combined pediatric and adult ECLS circuits using an i-cor pump head and either an i-cor console or Medos DeltaStream MDC console, a Medos Hilite 2400 LT oxygenator for the pediatric ECLS circuit, and a Medos Hilite 7000 LT oxygenator for the adult ECLS circuit. The circuit was primed with lactated Ringer's solution and human packed red blood cells (hematocrit 40%). Trials were conducted at various flow rates (pediatric circuit: 0.5 and 1L/min; adult circuit: 2 and 4L/min) under nonpulsatile and pulsatile modes (pulsatile amplitude: 1000-5000rpm [1000 rpm increments] for i-cor pump, 500-2500rpm [500 rpm increments] for Medos pump) at 36°C. In an additional protocol, fresh whole blood was used to test hemolysis under nonpulsatile and pulsatile modes using the two pump systems in adult ECLS circuits. Under pulsatile mode, energy equivalent pressures (EEP) were always greater than mean pressures for the two systems. Total hemodynamic energy (THE) and surplus hemodynamic energy (SHE) levels delivered to the patient increased with increasing pulsatile amplitude and decreased with increasing flow rate. The i-cor pump outperformed at low flow rates, but the Medos pump performed superiorly at high flow rates. There was no significant difference between two pumps in percentage of THE loss. The plasma free hemoglobin level was always higher in the Medos DP3 pulsatile group at 4 L/min compared to others. Pulsatile control algorithms of Medos and i-cor consoles had great effects on pulsatility. Although high pulsatile amplitudes delivered higher levels of hemodynamic energy to the patient, the high rotational speeds increased the risk of hemolysis. Use of proper pulsatile amplitude settings and intermittent pulsatile mode are suggested to achieve better pulsatility and decrease the risk of hemolysis. Further optimized pulsatile control algorithms are needed.© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…