JACC. Heart failure · Dec 2014
Randomized Controlled Trial Multicenter Study Comparative StudyReduced risk for inappropriate implantable cardioverter-defibrillator shocks with dual-chamber therapy compared with single-chamber therapy: results of the randomized OPTION study.
- Christof Kolb, Marcio Sturmer, Peter Sick, Sebastian Reif, Jean Marc Davy, Giulio Molon, Jörg Otto Schwab, Giuseppe Mantovani, Dan Dan, Carsten Lennerz, Alberto Borri-Brunetto, and Dominique Babuty.
- Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Faculty of Medicine, Technische Universität München, Munich, Germany. Electronic address: kolb@dhm.mhn.de.
- JACC Heart Fail. 2014 Dec 1; 2 (6): 611-9.
ObjectivesThe OPTION (Optimal Anti-Tachycardia Therapy in Implantable Cardioverter-Defibrillator Patients Without Pacing Indications) trial sought to compare long-term rates of inappropriate shocks, mortality, and morbidity between dual-chamber and single-chamber settings in implantable cardioverter-defibrillators (ICDs) patients.BackgroundThe use of dual-chamber ICDs potentially allows better discrimination of supraventricular arrhythmias and thereby reduces inappropriate shocks. However, it may lead to detrimental ventricular pacing.MethodsThis prospective multicenter, single-blinded trial enrolled 462 patients with de novo primary or secondary prevention indications for ICD placement and with left ventricular ejection fractions ≤40% despite optimal tolerated pharmacotherapy. All patients received atrial leads and dual-chamber defibrillators that were randomized to be programmed either with dual-chamber or single-chamber settings. In the dual-chamber setting arm, the PARAD+ algorithm, which differentiates supraventricular from ventricular arrhythmias, and SafeR mode, to minimize ventricular pacing, were activated. In the single-chamber setting arm, the acceleration, stability, and long cycle search discrimination criteria were activated, and pacing was set to VVI 40 beats/min. Ventricular tachycardia detection was required at rates between 170 and 200 beats/min, and ventricular fibrillation detection was activated above 200 beats/min.ResultsDuring a follow-up period of 27 months, the time to the first inappropriate shock was significantly longer in the dual-chamber setting arm (p = 0.012, log-rank test), and 4.3% of patients in the dual-chamber setting group compared with 10.3% in the single-chamber setting group experienced inappropriate shocks (p = 0.015). Rates of all-cause death or cardiovascular hospitalization were 20% for the dual-chamber setting group and 22.4% for the single-chamber setting group and satisfied the pre-defined margin for equivalence (p < 0.001).ConclusionsTherapy with dual-chamber settings for ICD discrimination combined with algorithms for minimizing ventricular pacing was associated with reduced risk for inappropriate shock compared with single-chamber settings, without increases in mortality and morbidity. (Optimal Anti-Tachycardia Therapy in Implantable Cardioverter-Defibrillator [ICD] Patients Without Pacing Indications [OPTION]; NCT00729703).Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.