• Beijing Da Xue Xue Bao · Oct 2017

    [Comparison between continuous noninvasive hemoglobin monitoring and venous blood hemoglobin monitoring in children with kidney disease].

    • X Y Liu, X Q Yang, H J Xiao, and J Ding.
    • Department of Pediatrics, Peking University First Hospital, Beijing 100034, China.
    • Beijing Da Xue Xue Bao. 2017 Oct 18; 49 (5): 778-782.

    ObjectiveTo assess the correlation and consistency between continuous noninvasive hemoglobin detection and venous blood hemoglobin detection in children with kidney disease, and try to analyze the affecting factors. Try to provide a reference for the monitoring of hemoglobin in children with kidney disease by continuous noninvasive hemoglobin detection technique.MethodsEighty-five inpatient children with kidney disease, 50 boys (58.8%) and 35 girls (41.2%), aged from 3 years old to 18 years old (9.35±4.29) were included finally. Noninvasive hemoglobin monitoring by spectrophotometry (SpHb) was stably read by PRONTO-7, selecting the ring finger of the non handedness as the detection site. And then the venous blood hemoglobin of the same patient was collected in 5 minutes as the true hemoglobin (tHb) by Beckman coulter DXH-800. The data of SpHb and tHb were compared and analyzed using SPSS 17.0 and MedCalc.ResultsCorrelation analysis showed data of SpHb and tHb were with significant correlation, the correlation coefficient between SpHb and tHb was 0.85 (P<0.05). Bland-Altman plot points suggested that the mean of differences between SpHb and tHb was -1.3 g/dL. The 95% CI of agreement of SpHb-tHb was -4.2-1.5 g/dL, suggesting that the average measurement result of SpHb was lower than that of tHb. The mean of differences as percent between SpHb and tHb was -9.8%, 95% CI of agreement was (-35.9%, 16.2%) exceeding the acceptable range of true value ±6%. The consistent rate of non-invasive hemoglobin detection and venous blood hemoglobin detection was 31.8%, the 95% CI of consistent rate was (21.7%, 41.9%). The chi square test of the fourfold table showed that the diagnosis of anemia with SpHb was of high sensitivity, but the specificity was low, the false positive rate was high, and the difference was statistically significant (P=0.000).ConclusionThere was a significant correlation between SpHb and tHb in the children with kidney disease. Noninvasive hemoglobin measurement can be used for monitoring of changes of hemoglobin in children with kidney diseases. But the consistency between SpHb and tHb needs to be improved. Noninvasive hemoglobin measurement could not replace the venous hemoglobin measurement. It could not be used for the diagnosis of anemia, and the accuracy of hemoglobin concentration measurement in children with kidney disease should be further explored.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.