• Ann Occup Hyg · Aug 2013

    Flow and containment characteristics of a sash-less, variable-height inclined air-curtain fume hood.

    • Rong Fung Huang, Jia-Kun Chen, and Wei-Lun Hung.
    • Department of Mechanical Engineering, National Taiwan University of Science and Technology, Taipei, Republic of China. rfhuang@mail.ntust.edu.tw
    • Ann Occup Hyg. 2013 Aug 1; 57 (7): 934-52.

    AbstractTo increase containment efficiency and reduce energy consumption, a sash-less, variable-height inclined air-curtain fume hood (sIAC hood) was developed and tested by a laser-assisted flow visualization technique and tracer-gas detection method. This novel design requires neither sash nor baffle. The sIAC hood employed the inclined push-pull air-curtain technique and two deflection plates installed on the side walls of the hood to induce a tetra-vortex flow structure. The results of flow visualization showed that the slot for suction flow, offset from the slot for the up-blowing jet, caused the air curtain to incline towards the rear wall, thus enhancing the robustness of the tetra-vortex flow structure. Such a flow structure could reduce the influence of draught and human walk-by across the hood face. The containment around the central area of the hood was isolated by the inclined push-pull air curtain. The pollutants carried by the reverse flow induced by the flow separation were guided by the deflection plates from the side walls towards the rear, thus contributing to the formation of the tetra-vortex flow structure. The up/down movable ceiling positioned the suction slot close to the device's pollutant emission opening, but left room (less than 50 cm) for unrestricted hand movement. Testing was carried out based on the methodology described in EN14175. The results of a static test showed that small face velocities of 0.25 and 0.16 m s(-1) were enough to obtain nearly null leakage levels for low and tall pollutant sources. The results of a traversing plate test showed that the face velocity, 0.32 m s(-1), would cause negligibly small leakage levels. The sIAC hood could obtain significantly higher containment efficiency than a conventional hood by operating at a face velocity significantly lower than that of conventional hoods.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?

    User can't be blank.

    Content can't be blank.

    Content is too short (minimum is 15 characters).

    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.