    A dual modality approach to quantitative quality control in emission tomography.

    • Michael D R Thomas, Dale L Bailey, and Lefteris Livieratos.
    • Joint Department of Physics, Institute of Cancer Research, Royal Marsden Hospital, Sutton, UK.
    • Phys Med Biol. 2005 Aug 7; 50 (15): N187-94.

    AbstractRoutine quality control (QC) and optimization of image quality of reconstructed images in single photon emission computed tomography (SPECT) and positron emission tomography (PET) remains a relatively qualitative exercise. With the advent of combined SPECT/CT and PET/CT devices, and accurate post hoc co-registration algorithms, the potential exists to utilize high resolution structural information for QC evaluation in addition to their use for anatomical correlation in clinical studies. The aim of this work was to explore, in principle, the uses of x-ray CT data of QC phantoms used in SPECT and PET to develop more objective assessments of performance of the emission tomographic (ET) devices and reconstructed data. A CT reconstruction of a novel ET QC phantom was segmented into the various compartments it contained. Using software, the voxel values in the different compartments were then altered to correspond to the concentration of the radioactivity in the actual scan of the same phantom on the SPECT system. This produces a high resolution version of a 'perfect' ET scan. Image co-registration techniques were then used to spatially align the synthetic high resolution SPECT scan to the measured SPECT scan. Various parameters can then be objectively derived from the registered data, for example, image contrast, spatial resolution, spatial non-uniformity, etc. In this study, we have used this approach to estimate spatial resolution (full width at half maximum, FWHM) and recovered contrast in reconstructed images of a SPECT phantom. Two independent methods were used to measure spatial resolution, obtaining excellent agreement. In conclusion, the ability to produce high resolution synthetic phantoms in emission tomography QC affords an objective approach to assessing system performance and optimizing protocols which is readily automated and quantifiable.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,706,642 articles already indexed!

We guarantee your privacy. Your email address will not be shared.