J Comput Assist Tomogr · Jan 1998
Comparative Study Clinical Trial Controlled Clinical TrialImaging articular cartilage defects with 3D fat-suppressed echo planar imaging: comparison with conventional 3D fat-suppressed gradient echo sequence and correlation with histology.
- S Trattnig, M Huber, M J Breitenseher, H J Trnka, T Rand, A Kaider, T Helbich, H Imhof, and D Resnick.
- Department of Radiology, University of Vienna, Austria.
- J Comput Assist Tomogr. 1998 Jan 1; 22 (1): 8-14.
PurposeOur goal was to shorten examination time in articular cartilage imaging by use of a recently developed 3D multishot echo planar imaging (EPI) sequence with fat suppression (FS). We performed comparisons with 3D FS GE sequence using histology as the standard of reference.MethodTwenty patients with severe gonarthrosis who were scheduled for total knee replacement underwent MRI prior to surgery. Hyaline cartilage was imaged with a 3D FS EPI and a 3D FS GE sequence. Signal intensities of articular structures were measured, and contrast-to-noise (C/N) ratios were calculated. Each knee was subdivided into 10 cartilage surfaces. From a total of 188 (3D EPI sequence) and 198 (3D GE sequence) cartilage surfaces, 73 and 79 histologic specimens could be obtained and analyzed. MR grading of cartilage lesions on both sequences was based on a five grade classification scheme and compared with histologic grading.ResultsThe 3D FS EPI sequence provided a high C/N ratio between cartilage and subchondral bone similar to that of the 3D FS GE sequence. The C/N ratio between cartilage and effusion was significantly lower on the 3D EPI sequence due to higher signal intensity of fluid. MR grading of cartilage abnormalities using 3D FS EPI and 3D GE sequence correlated well with histologic grading. 3D FS EPI sequence agreed within one grade in 69 of 73 (94.5%) histologically proven cartilage lesions and 3D FS GE sequence agreed within one grade in 76 of 79 (96.2%) lesions. The gradings were identical in 38 of 73 (52.1%) and in 46 of 79 (58.3%) cases, respectively. The difference between the sensitivities was statistically not significant.ConclusionThe 3D FS EPI sequence is comparable with the 3D FS GE sequence in the noninvasive evaluation of advanced cartilage abnormalities but reduces scan time by a factor of 4.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.