• Bull Hosp Jt Dis (2013) · Dec 2015

    Comparative Study

    The Impact of Posterior Wear on Reverse Shoulder Glenoid Fixation.

    • Richard Friedman, Nicholas Stroud, Kaycee Glattke, Pierre-Henri Flurin, Thomas W Wright, Joseph D Zuckerman, and Christopher P Roche.
    • Bull Hosp Jt Dis (2013). 2015 Dec 1; 73 Suppl 1: S15-20.

    IntroductionAchieving glenoid fixation with posterior bone loss can be challenging. The purpose of this study was to quantify the impact of two different sizes of posterior glenoid defects (10° and 20°) on reverse shoulder arthroplasty (rTSA) glenoid baseplate fixation and determine if utilizing different sizes of posterior augmented baseplates (8° and 16°) with off-axis reaming provides comparable fixation to using a standard baseplate with different amounts of eccentric reaming.MethodsWe quantified the impact of 10° and 20° posterior glenoid defects on rTSA baseplate fixation in composite scapulae using the ASTM F2028-14 rTSA glenoid loosening test method. Forty-two total implants (N = 7 for each size defect and for each type of baseplate) were tested at 750 N for 10,000 cycles. Baseplate displacement was measured before and after cyclic loading in the superior-inferior and anterior-posterior directions. Statistical analysis was performed with a two-tailed unpaired Student's t-test (significance defined as p < 0.05) to compare prosthesis displacements relative to each scapula (10° and 20° posterior defects for each type of baseplate versus the non-defect control) before and after cyclic loading.ResultsAll glenoid baseplates remained well-fixed after cyclic loading in composite scapulae without a defect and in scapulae with posterior defects. Increased pre- and post-cyclic displacement was observed with increased posterior defect size and differences in displacement were observed between standard and augmented baseplates. Augmented baseplates were observed to remove significantly less bone than standard baseplates when correcting posterior defects, regardless of size.DiscussionBoth standard baseplates with eccentric reaming and two different sizes of augmented baseplates with off-axis reaming successfully maintained fixation following cyclic loading in composite scapula with corrected 10° and 20° posterior glenoid defects. Augmented glenoids may be more advantageous long-term from a fixation perspective as they preserve more subchondral glenoid bone due to the minimal reaming occurring by the off-axis method. Mid and long-term clinical follow-up comparisons of outcomes are necessary between these two techniques.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…