    Volume-controlled ventilation is made possible in infants by using compliant breathing circuits with large compression volume.

    • M Badgwell, J Swan, and A C Foster.
    • Department of Anesthesiology, Texas Tech University Health Sciences Center, Lubbock 79430, USA.
    • Anesth. Analg. 1996 Apr 1; 82 (4): 719-23.

    AbstractWe studied the weight dependency of set tidal volume (VTset) during volume-controlled ventilation of 80 infants (ASA physical status I-IV, 0.7-20 kg), including prematures, neonates, and exprematures, who were anesthetized for major and minor surgery, including abdominal, thoracic, and neurosurgical procedures. After neuromuscular blockade and endotracheal intubation, infant's lungs were ventilated with an Ohmeda 7800 volume-limited ventilator and either a pediatric or adult circle breathing system (PC or AC) or a Bain circuit (Ba) and a pediatric- or adult-sized bellows (PB or AB). Except for larger and older infants in the ACAB group, body weight, age, peak inspiratory pressure (PIP), ETCO2, and SPO2 did not differ among groups. Compression volume loss was large in the five circuits tested. We found that VTset/kg varied with weight in a curvilinear relationship where y represents volume added and x represents PIP (y = 175.02x-0.87; r2 - 0.87), whereby VTset is approximately 150-200 mL/kg for a 1-kg infant and approximately 25 mL/kg for infants > or = 10 kg. Ventilation was adequate in each infant, except for one with extremely poor pulmonary compliance. We conclude that large compression volumes associated with compliant breathing systems make possible the use of volume-controlled ventilators in small infants.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.