    Combining prospective motion correction and distortion correction for EPI: towards a comprehensive correction of motion and susceptibility-induced artifacts.

    • Rainer Boegle, Julian Maclaren, and Maxim Zaitsev.
    • Department of Radiology Medical Physics, University Hospital Freiburg, Freiburg, Germany. Rainer.Boegle@googlemail.com
    • MAGMA. 2010 Sep 1; 23 (4): 263-73.

    ObjectState-of-the-art MR techniques that rely on echo planar imaging (EPI), such as real-time fMRI, are limited in their applicability by both subject motion and B(0) field inhomogeneities. The goal of this work is to demonstrate that in principle it is possible to accurately predict the B(0) field inhomogeneities that occur during echo planar imaging in the presence of large scale head motion and apply this knowledge for distortion correction.Materials And MethodsIn this work, prospective motion correction is combined with a field-prediction method and a method for correcting geometric distortions in EPI. To validate the methods, echo planar images were acquired of a custom-made phantom rotated to different angles relative to the B(0) field. For each orientation, field maps were acquired for comparison with the field predictions.ResultsThe calculated field maps are very similar to the measured field maps for all orientations used in the experiments. The root mean squared error (RMSE) of the difference maps was between 15 to 20 Hz. The quality of distortion correction using calculated field maps is comparable to distortion correction done with measured field maps.ConclusionThe results suggest that distortion-free echo planar imaging of moving objects may be feasible if prospective motion correction is combined with a field inhomogeneity estimation approach.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?

    User can't be blank.

    Content can't be blank.

    Content is too short (minimum is 15 characters).

    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.