• Magn Reson Imaging · Nov 2018

    Multicenter Study

    Scan-rescan repeatability and cross-scanner comparability of DTI metrics in healthy subjects in the SPRINT-MS multicenter trial.

    • Xiaopeng Zhou, Ken E Sakaie, Josef P Debbins, Sridar Narayanan, Robert J Fox, and Mark J Lowe.
    • School of Health Sciences, Purdue University, 550 Stadium Mall Drive, West Lafayette, IN 47907, USA.
    • Magn Reson Imaging. 2018 Nov 1; 53: 105-111.

    PurposeTo assess intrascanner repeatability and cross-scanner comparability for diffusion tensor imaging (DTI) metrics in a multicenter clinical trial.MethodsDTI metrics (including longitudinal diffusivity [LD], fractional anisotropy [FA], mean diffusivity [MD], and transverse diffusivity [TD]) from pyramidal tracts for healthy controls were calculated from images acquired on twenty-seven 3T MR scanners (Siemens and GE) with 6 different scanner models and 7 different software versions as part of the NN102/SPRINT-MS clinical trial. Each volunteer underwent two scanning sessions on the same scanner. Signal-to-noise ratio (SNR) and signal-to-noise floor ratio (SNFR) were also assessed.ResultsDTI metrics showed good scan-rescan repeatability. There were no significant differences between scans and rescans in LD, FA, MD, or TD values. Although the cross-scanner coefficient of variation (CV) values for all DTI metrics were <5.7%, significant differences were observed for LD (p < 3.3e-5) and FA (p < 0.0024) when GE scanners were compared with Siemens scanners. Significant differences were also observed for SNR when comparing GE scanners and Siemens Skyra scanners (p < 1.4e-7) and when comparing Siemens Skyra scanners and TIM Trio scanners (p < 1.0e-10). Analysis of background signal also demonstrated differences between GE and Siemens scanners in terms of signal statistics. The measured signal intensity from a background noise region of interest was significantly higher for GE scanners than for Siemens scanners (p < 1.2e-12). Significant differences were also observed for SNFR when comparing GE scanners and Siemens Skyra scanners (p < 2.5e-11), GE scanners and Siemens Trio scanners (p < 7.5e-11), and Siemens Skyra scanners and TIM Trio scanners (p < 2.5e-9).ConclusionsThe good repeatability of the DTI metrics among the 27 scanners used in this study confirms the feasibility of combining DTI data from multiple centers using high angular resolution sequences. Our observations support the feasibility of longitudinal multicenter clinical trials using DTI outcome measures. The noise floor level and SNFR are important parameters that must be assessed when comparing studies that used different scanner models.Copyright © 2018 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…