• Eur Spine J · Nov 2021

    Accuracy assessment of pedicle screw insertion with patient specific 3D‑printed guides through superimpose CT-analysis in thoracolumbar spinal deformity surgery.

    • J Cool, J van Schuppen, M A de Boer, and B J van Royen.
    • Department of Orthopedic Surgery, Amsterdam UMC, University of Amsterdam and Vrije Universiteit, Amsterdam Movement Sciences, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands.
    • Eur Spine J. 2021 Nov 1; 30 (11): 3216-3224.

    PurposeIn order to avoid pedicle screw misplacement in posterior spinal deformity surgery, patient specific 3D‑printed guides can be used. An accuracy assessment of pedicle screw insertion can be obtained by superimposing CT-scan images from a preoperative plan over those of the postoperative result. The aim of this study is to report on the accuracy of drill guide assisted pedicle screw placement in thoracolumbar spinal deformity surgery by means of a superimpose CT-analysis.MethodsConcomitant with the clinical introduction of a new technique for drill guide assisted pedicle screw placement, the accuracy of pedicle screw insertion was analyzed in the first patients treated with this technique by using superimpose CT-analysis. Deviation from the planned ideal intrapedicular screw trajectory was classified according to the Gertzbein scale.ResultsSuperimpose CT-analysis of 99 pedicle screws in 5 patients was performed. The mean linear deviation was 0.92 mm, the mean angular deviation was 2.92° with respect to the preoperatively planned pedicle screw trajectories. According to the Gertzbein scale, 100% of screws were found to be positioned within the "safe zone".ConclusionThe evaluated patient specific 3D-printed guide technology was demonstrated to constitute a safe and accurate tool for precise pedicle screw insertion in spinal deformity surgeries. Superimpose CT-analysis showed a 100% accuracy of pedicle screw placement without any violation of the pedicle wall or other relevant structures. We recommend a superimpose CT-analysis for the first consecutive patients when introducing new technologies into daily clinical practice, such as intraoperative imaging, navigation or robotics.© 2021. The Author(s).

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…