• Critical care medicine · Nov 2013

    Assessment of Dynamic Mechanical Properties of the Respiratory System During High-Frequency Oscillatory Ventilation.

    • Raffaele L Dellacà, Emanuela Zannin, Maria L Ventura, Giulio Sancini, Antonio Pedotti, Paolo Tagliabue, and Giuseppe Miserocchi.
    • 1TBM Lab, Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano University, Milano, Italy. 2Neonatal Intensive Care Unit, MBBM Foundation, Monza, Italy. 3Department of Health Sciences, Università di Milano Bicocca, Monza, Italy.
    • Crit. Care Med.. 2013 Nov 1;41(11):2502-11.

    Objectives1) To investigate the possibility of estimating respiratory system impedance (Zrs, forced oscillation technique) by using high-amplitude pressure oscillations delivered during high-frequency oscillatory ventilation; 2) to characterize the relationship between Zrs and continuous distending pressure during an increasing/decreasing continuous distending pressure trial; 3) to evaluate how the optimal continuous distending pressure identified by Zrs relates to the point of maximal curvature of the deflation limb of the quasi-static pressure-volume curve.DesignProspective laboratory animal investigation.SettingExperimental medicine laboratory.SubjectsEight New Zealand rabbits.InterventionsThe rabbits were ventilated with high-frequency oscillatory ventilation. Zrs was measured while continuous distending pressure was increased and decreased between 2 and 26 cm H2O in 1-minute steps of 4 cm H2O. At each step, a low-amplitude (6 cm H2O) sinusoidal signal was alternated with a high-amplitude (18 cm H2O) asymmetric high-frequency oscillatory ventilation square pressure waveform. Pressure-volume curves were determined at the end of the continuous distending pressure trial. All measurements were repeated after bronchoalveolar lavage.Measurements And Main ResultsZrs was estimated from flow and pressure measured at the inlet of the tracheal tube and expressed as resistance (Rrs) and reactance (Xrs). Linear correlation between the values, measured by applying the small-amplitude sinusoidal signal and the ventilator waveform, was good for Xrs (r = 0.95 ± 0.04) but not for Rrs (r = 0.60 ± 0.34). Following lavage, the Xrs-continuous distending pressure curves presented a maximum on the deflation limb, identifying an optimal continuous distending pressure that was, on average, 1.1 ± 1.7 cm H2O below the point of maximal curvature of the deflation limb of the pressure-volume curves.ConclusionsXrs can be accurately measured during high-frequency oscillatory ventilation without interrupting ventilation and/or connecting additional devices. An optimal continuous distending pressure close to the point of maximal curvature of the deflation limb of quasi-static pressure-volume curve can be identified by measuring Zrs during a decreasing continuous distending pressure trial. Zrs might constitute a useful bedside tool for monitoring lung mechanics and improving the continuous distending pressure optimization during high-frequency oscillatory ventilation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…