    Gastric emptying measurement of liquid nutrients using the (13)C-octanoate breath test in critically ill patients: a comparison with scintigraphy.

    • Marianne Chapman, Carly M Burgstad, Nam Q Nguyen, Richard H Holloway, Laura K Bryant, Adam Deane, Max Bellon, and Dylan Bartholomeuz.
    • Department of Gastroenterology and Hepatology, Royal Adelaide Hospital, North Terrace, Adelaide, South Australia 5000, Australia. quoc.nguyen@health.sa.gov.au
    • Intensive Care Med. 2013 Jul 1;39(7):1238-46.

    PurposeScintigraphy is considered the most accurate technique for the measurement of gastric emptying (GE) but, for patients in the intensive care unit, it is technically demanding, involves radiation and can interfere with care. The (13)C-octanoate breath test ((13)C-OBT) is a simple, non-invasive technique that does not involve radiation exposure.AimTo evaluate the performance of the (13)C-OBT in the assessment of GE in critically ill patients.MethodsThe GE was assessed in 33 mechanically ventilated patients (23 M; 54.3 ± 3.0 yrs; APACHE II: 22.0 ± 1.1). Following test meal administration (100 ml Ensure(®)), concurrent scintigraphic measurement and breath samples ((13)C-OBT) were collected over 4 h. Scintigraphic meal retention was determined and the gastric emptying coefficient (GEC) and half emptying time [t50(BT)] were calculated for the (13)C-OBT. Delayed GE was defined as meal retention >13 % at 180 min.ResultsDelayed GE was identified in 27/33 patients. Meal retention correlated modestly with t50(BT) (r = 0.55-0.66; P < 0.001) and well with GEC (r = -0.63 to -0.74; P < 0.0001). The strength of agreement between the two techniques was highest between GEC and retention at 120 min. The best cut-off GEC for defining delayed GE was 3.25 (AUC = 0.75; 95 % CI = 0.52-0.99; P = 0.05), with 89 % sensitivity and 67 % specificity to detect delayed GE. The GE was delayed in all (23/23) patients with feed intolerance (GRV > 250 ml) on scintigraphy and 91 % (21/23) patients on (13)C-OBT.ConclusionIn critical illness, there was a correlation between (13)C-OBT and gastric scintigraphy, with GEC performing as a better and more sensitive marker of detecting delayed GE than t50. However the relatively wide 95 % confidence intervals suggest that (13)C-OBT is more suitable as a technique to assess GE in a group setting for research studies rather than for individual patients in clinical practice.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…