• Br J Surg · Nov 2000

    Compliance properties of conduits used in vascular reconstruction.

    • N R Tai, H J Salacinski, A Edwards, G Hamilton, and A M Seifalian.
    • University Department of Surgery, Royal Free and University College Medical School, University College London and Royal Free Hospital, UK.
    • Br J Surg. 2000 Nov 1; 87 (11): 1516-24.

    BackgroundCompliance mismatch between native artery and prosthetic graft used for infrainguinal bypass is implicated in the aetiology of graft failure. The aim was to quantify the elastic properties of a new compliant poly(carbonate)polyurethane (CPU) vascular graft, and to compare the compliance properties of grafts made from CPU, expanded polytetrafluoroethylene (ePTFE), Dacron and human saphenous vein with that of human muscular artery.MethodsA pulsatile flow phantom was used to perfuse vessel and prosthetic graft segments at physiological pulse pressure and flow. Intraluminal pressure was measured using a Millar Mikro-tip catheter transducer and vessel wall motion was determined with duplex ultrasonography using an echo-locked wall-tracking system. Diametrical compliance and a stiffness index were then calculated for each type of conduit over mean pressures ranging from 30 to 100 mmHg by 10-mmHg increments.ResultsThe compliance values of CPU and artery (mean over the pressure range) were similar (mean(s.d.) 8.1(0.4) and 8.0(5.9) per cent per mmHg x 10(-2) respectively), although the elastic behaviour of artery was anisotropic unlike CPU, which was isotropic. Dacron and ePTFE grafts had lower compliance values (1.8(1.2) and 1.2(0.3)per cent per mmHg x 10(-2) respectively, averaged over the pressure range). In both these cases, compliance and stiffness differed significantly from that of artery over a mean pressure range of 30-90 mmHg. Human saphenous vein exhibited anisotropic behaviour and, although compliant at low pressure (30 mmHg), was markedly incompliant at higher pressures.ConclusionCompliant polyurethane grafts offer a greater degree of compliance match than either ePTFE or Dacron.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…