    Potentially harmful effects of inspiratory synchronization during pressure preset ventilation.

    • J C M Richard, A Lyazidi, E Akoumianaki, S Mortaza, R L Cordioli, J C Lefebvre, N Rey, L Piquilloud, G F Sferrazza Papa, G F Sferrazza-Papa, A Mercat, and L Brochard.
    • Intensive Care Unit, University Hospital of Geneva, 4 Rue Gabrielle Perret-Gentil, 1211, Genève 14, Switzerland, jcmb.richard@gmail.com.
    • Intensive Care Med. 2013 Nov 1; 39 (11): 2003-10.

    PurposePressure preset ventilation (PPV) modes with set inspiratory time can be classified according to their ability to synchronize pressure delivery with patient's inspiratory efforts (i-synchronization). Non-i-synchronized (like airway pressure release ventilation, APRV), partially i-synchronized (like biphasic airway pressure), and fully i-synchronized modes (like assist-pressure control) can be distinguished. Under identical ventilatory settings across PPV modes, the degree of i-synchronization may affect tidal volume (VT), transpulmonary pressure (PTP), and their variability. We performed bench and clinical studies.MethodsIn the bench study, all the PPV modes of five ventilators were tested with an active lung simulator. Spontaneous efforts of -10 cmH2O at rates of 20 and 30 breaths/min were simulated. Ventilator settings were high pressure 30 cmH2O, positive end-expiratory pressure (PEEP) 15 cmH2O, frequency 15 breaths/min, and inspiratory to expiratory ratios (I:E) 1:3 and 3:1. In the clinical studies, data from eight intubated patients suffering from acute respiratory distress syndrome (ARDS) and ventilated with APRV were compared to the bench tests. In four additional ARDS patients, each of the PPV modes was compared.ResultsAs the degree of i-synchronization among the different PPV modes increased, mean VT and PTP swings markedly increased while breathing variability decreased. This was consistent with clinical comparison in four ARDS patients. Observational results in eight ARDS patients show low VT and a high variability with APRV.ConclusionDespite identical ventilator settings, the different PPV modes lead to substantial differences in VT, PTP, and breathing variability in the presence spontaneous efforts. Clinicians should be aware of the possible harmful effects of i-synchronization especially when high VT is undesirable.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…