• Respiratory care · Nov 2023

    A Comparison of Proximal and Tracheal Airway Pressures During Pressure Controlled Ventilation.

    • Mark O Zander, Nikola Stankovic, Mirko Meboldt, Thomas O Erb, Jürg Hammer, and Marianne Schmid Daners.
    • Product Development Group Zurich, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland.
    • Respir Care. 2023 Nov 25; 68 (12): 163916451639-1645.

    BackgroundAirway pressure is usually measured by sensors placed in the ventilator or on the ventilator side of the endotracheal tube (ETT), at the Y-piece. These remote measurements serve as a surrogate for the tracheal or alveolar pressure. Tracheal pressure can only be predicted correctly by using a model that incorporates the pressure at the remote location, the flow through the ETT, and the resistance of the ETT if the latter is a predictable function of Y-piece flow. However, this is not consistently appropriate, and accuracy of prediction is hampered.MethodsThis in vitro study systematically examined the ventilator pressure in dependence of compliance of the respiratory system (CRS), inspiratory time, and expiratory time during pressure-controlled ventilation by using a small intratracheal pressure sensor and a mechanical lung simulator. Pressures were measured simultaneously at the ventilator outlet, at the Y-piece, and in the trachea during pressure-controlled ventilation with a peak inspiratory pressure of 20 cm H2O and a PEEP of 5 cm H2O while changing CRS (10, 30, 60, 90, and 100 mL/cm H2O) and varying inspiratory time and expiratory time.ResultsTracheal pressures were always lower (maximum 8 cm H2O during inspiration) or higher (maximum 4 cm H2O during expiration) than the pressures measured proximal to the ETT if zero-flow conditions were not achieved at the end of the breathing cycles.ConclusionsDependent on CRS and the breathing cycle, tracheal pressures deviated from those measured proximal to the ETT under non-zero-flow conditions. Intratracheal pressure and pressure curve dynamics can differ greatly from the ventilator pressure, depending on the ventilator setting and the CRS. The small pressure sensor may be used as a measurement method of tracheal pressure via integration onto an ETT.Copyright © 2023 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.