• J Clin Monit Comput · Feb 2024

    The 10 Hz dynamic response of a fluid-filled pressure monitoring system is a novel alternative to the fast flush test and indicative of unacceptable systolic pressure overshoot.

    • Tomoki Hirahata, Shuichi Hashimoto, Hiroaki Watanabe, Shin-Ichi Yagi, Mitsutaka Edanaga, and Michiaki Yamakage.
    • Department of Anesthesiology, School of Medicine, Sapporo Medical University, South-1, West-16, Chuo-ku, Sapporo, 060-8543, Japan. tomoki.hirahata@gmail.com.
    • J Clin Monit Comput. 2024 Feb 4.

    AbstractThe standard method for qualitatively evaluating the dynamic response is to see if the gain of the amplitude spectrum curve approaches 1 (input signal = output signal) over the frequency band of the blood pressure waveform. In a previous report, Watanabe reported that Gardner's natural frequency and damping coefficient, which are widely used as evaluation methods, do not reflect the dynamic response of the circuit. Therefore, new parameters for evaluating the dynamic response of pressure monitoring circuits were desired. In this study, arterial pressure catheters with length of 30, 60, 150, and 210 cm were prepared, and a blood pressure wave calibrator, two pressure monitors with analog output and a personal computer were used to analyze blood pressure monitoring circuits. All data collection and analytical processes were performed using step response analysis program. The gain at 10 Hz was close to 1 and the systolic blood pressure difference was small in the short circuits (30 cm, 60 cm), and the gain at 10 Hz was 1.3-1.5 in the 150 cm circuit and over 1.7 in the 210 cm circuit. The difference in systolic blood pressure increased in proportion to the length of the circuit. It could also be inferred that the gain at 10 Hz should be less than 1.2 to meet a clinically acceptable blood pressure difference. In conclusion, the gain at 10 Hz is sufficiently useful as an indicator to determine the correct systolic blood pressure.© 2024. The Author(s), under exclusive licence to Springer Nature B.V.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.