• Can J Anaesth · May 2024

    A threshold of 100 or more colony-forming units on the anesthesia machine predicts bacterial pathogen detection: a retrospective laboratory-based analysis.

    • Franklin Dexter, Kaitlin M Walker, Carmen Troncoso Brindeiro, Chase P Loftus, Cornelie C L Banguid, and Randy W Loftus.
    • University of Iowa, Iowa City, IA, USA. franklin-dexter@uiowa.edu.
    • Can J Anaesth. 2024 May 1; 71 (5): 600610600-610.

    PurposePreventing the spread of pathogens in the anesthesia work area reduces surgical site infections. Improved cleaning reduces the percentage of anesthesia machine samples with ≥ 100 colony-forming units (CFU) per surface area sampled. Targeting a threshold of < 100 CFU when cleaning anesthesia machines may be associated with a lower prevalence of bacterial pathogens. We hypothesized that anesthesia work area reservoir samples returning < 100 CFU would have a low (< 5%) prevalence of pathogens.MethodsIn this retrospective cohort study of bacterial count data from nine hospitals, obtained between 2017 and 2022, anesthesia attending and assistants' hands, patient skin sites (nares, axilla, and groin), and anesthesia machine (adjustable pressure-limiting valve and agent dials) reservoirs were sampled at case start and at case end. The patient intravenous stopcock set was sampled at case end. The isolation of ≥ 1 CFU of Staphylococcus aureus, methicillin-resistant Staphylococcus aureus, Enterococcus, vancomycin-resistant Enterococcus, gram-negative (i.e., Klebsiella, Acinetobacter, Pseudomonas, and Enterobacter spp.) or coagulase-negative Staphylococcus was compared for reservoir samples returning ≥ 100 CFU vs those returning < 100 CFU.ResultsBacterial pathogens were isolated from 24% (7,601/31,783) of reservoir samples, 93% (98/105) of operating rooms, and 83% (2,170/2,616) of cases. The ratio of total pathogen isolates to total CFU was < 0.0003%. Anesthesia machine reservoirs returned ≥ 100 CFU for 44% (2,262/5,150) of cases. Twenty-three percent of samples returning ≥ 100 CFU were positive for ≥ 1 bacterial pathogen (521/2,262; 99% lower confidence limit, 22%) vs 3% of samples returning < 100 CFU (96/2,888; 99% upper limit, 4%).ConclusionsAnesthesia machine reservoir samples returning < 100 CFU were associated with negligible pathogen detection. This threshold can be used for assessment of terminal, routine, and between-case cleaning of the anesthesia machine and equipment. Such feedback may be useful because the 44% prevalence of ≥ 100 CFU was comparable to the 46% (25/54) reported earlier from an unrelated hospital.© 2024. Canadian Anesthesiologists' Society.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…