    Performance of transport ventilator with patient-triggered ventilation.

    • E Miyoshi, Y Fujino, T Mashimo, and M Nishimura.
    • Intensive Care Unit, Osaka University Hospital, Osaka, Japan.
    • Chest. 2000 Oct 1;118(4):1109-15.

    ObjectivesTransport ventilators with inspiratory triggering functions and pressure support-control modes have recently become commercially available. We evaluated these ventilators in comparison with a standard ICU ventilator.Study DesignLaboratory study with a mechanical lung model.MethodsWe compared the performance of four transport ventilators (model 740, Mallinckrodt, Pleasanton, CA; TBird, Bird Products Corp, Palm Springs, CA; LTV1000, Pulmonetic Systems, Colton, CA; Esprit, Respironics, Vista, CA) with a standard ICU ventilator (model 7200ae; Mallinckrodt) using a test lung that simulated spontaneous breathing (compliance, 46.8 mL/cm H(2)O; resistance, 5 cm H(2)O/L/s). The settings of ventilators were positive end-expiratory pressure (PEEP) of 0 or 5 cm H(2)O, and pressure support (PS) of 0 or 10 cm H(2)O. The settings of the test lung were inspiratory time of 1 s, respiratory rate of 10/min, peak inspiratory flow of 40, 60, and 80 L/min. To evaluate inspiratory function at each setting, we measured the inspiratory delay time (DT), inspiratory trigger pressure (P-I), and the time for airway pressure to rise from the baseline pressure to 90% of the end-inspiratory pressure (T(90%)); for expiratory function, supraplateau expiratory pressure (P-E) and the time constant (taue) for pressure decrease during exhalation were evaluated. Oxygen requirement was assessed as the time required to empty a 3.5-L oxygen tank.ResultsFor inspiratory triggering, four transport ventilators had DT < 100 ms, which is considered clinically satisfactory, in all the settings except for PS 0 cm H(2)O, PEEP 0 cm H(2)O, and inspiratory flow of 80 L/min with LTV1000. P-I increased only in LTV1000 when PEEP was increased from 0 to 5 cm H(2)O. taue for the transport ventilators was > 50% shorter than for the ICU ventilator except for PS 0 cm H(2)O and PEEP 5 cm H(2)O with TBird. Oxygen requirement was lowest for the Esprit, followed by the 740, LTV1000, and TBird.ConclusionThe newer Food and Drug Administration-approved transport ventilators have performance indexes comparable to the ventilator currently used in ICUs and can probably be recommended for clinical use.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.