• Critical care medicine · Nov 2003

    Comparative Study

    Accuracy of automatic tube compensation in new-generation mechanical ventilators.

    • Serge Elsasser, Josef Guttmann, Reto Stocker, Georg Mols, Hans-Joachim Priebe, and Christoph Haberthür.
    • Department of Internal Medicine, Hospital Langenthal, Switzerland.
    • Crit. Care Med. 2003 Nov 1;31(11):2619-26.

    ObjectiveTo compare performance of flow-adapted compensation of endotracheal tube resistance (automatic tube compensation, ATC) between the original ATC system and ATC systems incorporated in commercially available ventilators.DesignBench study.SettingUniversity research laboratory.SubjectsThe original ATC system, Dräger Evita 2 prototype, Dräger Evita 4, Puritan-Bennett 840.InterventionsThe four ventilators under investigation were alternatively connected via different sized endotracheal tubes and an artificial trachea to an active lung model. Test conditions consisted of two ventilatory modes (ATC vs. continuous positive airway pressure), three different sized endotracheal tubes (inner diameter 7.0, 8.0, and 9.0 mm), two ventilatory rates (15/min and 30/min), and four levels of positive end-expiratory pressure (0, 5, 10, and 15 cm H2O).Measurements And Main ResultsPerformance of tube compensation was assessed by the amount of tube-related (additional) work of breathing (WOBadd), which was calculated on the basis of pressure gradient across the endotracheal tube. Compared with continuous positive airway pressure, ATC reduced inspiratory WOBadd by 58%, 68%, 50%, and 97% when using the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. Depending on endotracheal tube diameter and ventilatory pattern, inspiratory WOBadd was 0.12-5.2 J/L with the original ATC system, 1.5-28.9 J/L with the Puritan-Bennett 840, 10.4-21.0 J/L with the Evita 2 prototype, and 10.1-36.1 J/L with the Evita 4 (difference between each ventilator at identical test situations, p <.025). Expiratory WOBadd was reduced by 5%, 26%, 1%, and 70% with the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. The expiratory WOBadd caused by an endotracheal tube of 7.0 mm inner diameter was 5.5-42.2 J/L at a low ventilatory rate and 19.6-82.3 J/L at a high ventilatory rate. It was lowest with the original ATC system and highest with the Evita 4 ventilator (p <.025).ConclusionsFlow-adapted tube compensation by the original ATC system significantly reduced tube-related inspiratory and expiratory work of breathing. The commercially available ATC modes investigated here may be adequate for inspiratory but probably not for expiratory tube compensation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…