    Aerosol Delivery to Simulated Spontaneously Breathing Tracheostomized Adult Model With and Without Humidification.

    • Fai A Albuainain and Jie Li.
    • Division of Respiratory Care, Department of Cardiopulmonary Sciences, Rush University, Chicago, Illinois. Ms Albuainain is affiliated with the Department of Respiratory Care, Imam Abdulrahman Bin Faisal University, Jubail, Saudi Arabia.
    • Respir Care. 2024 Apr 23.

    BackgroundOptimal aerosol delivery methods for spontaneously breathing patients with a tracheostomy remain unclear. Thus, we aimed to assess the impact of nebulizer placement, flow settings, and interfaces on aerosol delivery by using a vibrating mesh nebulizer and a jet nebulizer in line with unheated humidification.MethodsAn 8.0-mm tracheostomy tube was connected to the lung model that simulates adult breathing parameters via a collecting filter. Albuterol sulfate (2.5 mg/3 mL) was administered via a vibrating mesh nebulizer and a jet nebulizer, which was placed in line with unheated humidification provided by a large-volume nebulizer, with FIO2 set at 0.28, with gas flows of 2 L/min versus 6 L/min. Nebulizers were placed in line distal and proximal to the lung model by using a tracheostomy collar and a T-piece. Conventional nebulization was tested using a vibrating mesh nebulizer and a jet nebulizer directly connected to the tracheostomy tube bypassing the humidification device. The drug was eluted from the collecting filter and assayed with ultraviolet spectrophotometry (276 nm).ResultsDuring in-line nebulizer placement with unheated humidification, the inhaled dose was 2-4 times higher with a gas flow of 2 L/min than 6 L/min, regardless of nebulizer type, placement, or interface (all P < .05). At 6 L/min, the inhaled dose was higher with proximal than distal placement when using both interfaces, but, at 2 L/min, the inhaled dose was lower with proximal placement. With a jet nebulizer, the tracheostomy collar generated a higher inhaled dose at proximal placement compared with the T-piece, whereas the T-piece resulted in a higher inhaled dose than the tracheostomy collar with distal placement, regardless of the flow settings. Compared with conventional nebulization using a vibrating mesh nebulizer, an in-line vibrating mesh nebulizer with a large-volume nebulizer at 2 L/min had a similar inhaled dose, regardless of nebulizer placement and interface. In contrast, the in-line jet nebulizer was influenced by both placement and interface.ConclusionsAerosol delivery with an in-line vibrating mesh nebulizer and jet nebulizer with unheated humidification was affected by nebulizer placement, interface, and gas flow settings.Copyright © 2024 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.