• Anesthesia and analgesia · Oct 1998

    Randomized Controlled Trial Comparative Study Clinical Trial

    A comparison of the disposable versus the reusable laryngeal mask airway in paralyzed adult patients.

    • J Brimacombe, C Keller, R Morris, and D Mecklem.
    • Department of Anaesthesia and Intensive Care, Cairns Base Hospital, Australia. 100236,2343@compuserve.com
    • Anesth. Analg. 1998 Oct 1;87(4):921-4.

    UnlabelledA disposable (polyvinyl chloride) laryngeal mask airway (LMA) with dimensions identical to, but physical properties different from (stiffer tube/thicker cuff), the reusable (silicone) LMA has recently become available. We performed a randomized, cross-over study of 60 paralyzed, anesthetized patients to test the hypothesis that the use of these devices was different in terms of ease of insertion, airway sealing pressure, fiberoptic position, and changes in intracuff pressure during N2O anesthesia. We also tested the hypothesis that the airway sealing pressure of the LMA is suboptimal if the cuff is inflated to a high intracuff pressure. Both the devices were inserted into each patient in random order, and their performance was assessed at two intracuff pressures (60 and 180 cm H2O) by a blind observer. Subsequently, intracuff pressures were measured during N2O anesthesia for the second device. Ease of insertion was similar: there was no difference in first attempt success rates (97% vs 98%) and insertion times (15 vs 13 s) for the disposable and reusable LMA, respectively. There were no differences in airway sealing pressure or fiberoptic position. Airway sealing pressure was significantly higher at 60 cm H2O intracuff pressure compared with the airway sealing pressure at 180 cm H2O for both devices (P < 0.02). During N2O anesthesia, the intracuff pressure remained stable for the disposable LMA but increased significantly for the reusable LMA. We conclude that the disposable and reusable LMAs perform similarly in paralyzed adult patients, but that the disposable LMA has more stable intracuff pressures during N2O anesthesia. Inflation of the LMA to high intracuff pressures produces a suboptimal seal.ImplicationsThis randomized, single-blind, within-patient study of 60 adult patients shows that the disposable (polyvinyl chloride) and reusable (silicone) laryngeal mask airways perform similarly, but that the disposable laryngeal mask airway has more stable intracuff pressures during N2O anesthesia. Inflation of either device to high intracuff pressures produces a suboptimal seal.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.