An evaluation of a new continuous blood pressure monitoring system in critically ill patients.
- S O Heard, A Lisbon, I Toth, and R Ramasubramanian.
- Department of Anesthesiology, UMass Memorial Medical Center, Worcester, MA 01655, USA. Stephen.Heard@umass.med.edu
- J Clin Anesth. 2000 Nov 1;12(7):509-18.
Study ObjectiveTo determine if the DxTek monitor, which is a device that measures blood pressure (BP) noninvasively and continuously by means of pulse velocity and wave shapes derived from the pulse oximeter optical plethysmograph and electrocardiogram is as accurate as an oscillometric cuff device when compared with intraarterial BP measurement.DesignProspective, comparative study.SettingUniversity Medical Center.Patients28 intensive care unit patients.InterventionsBlood pressures were reported every minute by intraarterial catheters and DxTek and every 10 minutes by an oscillometric monitor for 2 to 5 hours. DxTek calibration was performed initially and when specified patient manipulations by caretakers were performed (on average, every 100 minutes). Comparisons with intraarterial pressure included: 1) DxTek calibrated with arterial catheter pressure, 2) DxTek calibrated with oscillometric pressure, and 3) oscillometric pressure.Measurements And Main ResultsWhen comparing oscillometric pressure to intraarterial pressure, the averages of the mean differences (bias) were -4.0 mmHg for systolic (SBP) and < 1.5 mmHg for diastolic (DBP) and mean (MAP) pressures. The averages of the standard deviation of the differences (precisions) were 9.6, 6.4, and 6.3 mmHg, respectively. With the DxTek device calibrated to intraarterial pressure, comparison of the DxTek pressure to intraarterial pressure resulted in a bias < OR = 0.5 mmHg for all three pressures and an average precision of 10.1 mmHg for SBP, 6.0 mmHg for DBP, and 6.7 mmHg for MAP. With the DxTek device calibrated to the oscillometric pressure, the DxTek pressure compared to the intraarterial pressure resulted in average biases of -5.1, -0.8, and -2.2 mmHg and average precisions of 11.1, 7.7, and 8.1 mmHg for SBP, DBP, and MAP, respectively.ConclusionsThe DxTek monitor provides continuous, noninvasive BP measurements with an accuracy comparable to oscillometric devices.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.