Comparative Study
Electrical features of eighteen automated external defibrillators: A systematic evaluation.
- Fulvio Kette, Aldo Locatelli, Marcella Bozzola, Alberto Zoli, Yongqin Li, Marco Salmoiraghi, Giuseppe Ristagno, and Aida Andreassi.
- AREU Azienda Regionale Emergenza Urgenza (Lombardia Regional Emergency Service), Italy. Electronic address: f.kette@areu.lombardia.it.
- Resuscitation. 2013 Nov 1;84(11):1596-603.
AimAssessment and comparison of the electrical parameters (energy, current, first and second phase waveform duration) among eighteen AEDs.MethodEngineering bench tests for a descriptive systematic evaluation in commercially available AEDs. AEDs were tested through an ECG simulator, an impedance simulator, an oscilloscope and a measuring device detecting energy delivered, peak and average current, and duration of first and second phase of the biphasic waveforms. All tests were performed at the engineering facility of the Lombardia Regional Emergency Service (AREU).ResultsLarge variations in the energy delivered at the first shock were observed. The trend of current highlighted a progressive decline concurrent with the increases of impedance. First and second phase duration varied substantially among the AEDs using the exponential biphasic waveform, unlike rectilinear waveform AEDs in which phase duration remained relatively constant.ConclusionsThere is a large variability in the electrical features of the AEDs tested. Energy is likely not to be the best indicator for strength dose selection. Current and shock duration should be both considered when approaching the technical features of AEDs. These findings may prompt further investigations to define the optimal current and duration of the shock waves to increase the success rate in the clinical setting.Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.