• Br J Anaesth · Apr 2009

    Multicenter Study Comparative Study

    Comparison of two versions of the Vigileo-FloTrac system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements.

    • R Chatti, S de Rudniki, S Marqué, A S Dumenil, A Descorps-Declère, A Cariou, J Duranteau, M Aout, E Vicaut, and B P Cholley.
    • Département d'Anesthésie-Réanimation, AP-HP, Hôpital Lariboisière, Paris, France.
    • Br J Anaesth. 2009 Apr 1;102(4):463-9.

    BackgroundOur aim was to evaluate the validity of stroke volume measurements obtained using the Vigileo-FloTrac system in comparison with those obtained using oesophageal Doppler considered as a reference.MethodsProspective, multicentre study (four university hospitals), in which investigators were blinded to stroke volume values acquired simultaneously with the other technique. Two different versions of the Vigileo software (1.03 and 1.07) were studied and compared over two consecutive periods of time. Forty critically ill patients (three ICUs) and 20 high-risk surgical patients (one operating theatre) were studied over a 6-month period.ResultsTwo hundred and forty paired stroke volume values obtained using the second version of the Vigileo (1.07) yielded better correlation and agreement (R=0.48, P<0.001; bias=4 ml, limits of agreement: +/- 41 ml) than the 207 paired values obtained using version 1.03 (R=0.12, P=0.1; bias=1 ml, limits of agreement: +/- 75 ml). However, even with the second version, the percentage error in stroke volume measurement was 58%, a value still above the range considered clinically acceptable (30%).ConclusionsThe precision of stroke volume estimation using Vigileo-FloTrac has improved with the second version of the software (1.07), but remains insufficient to allow the replacement of the reference technique in the population studied.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.