• Anesth Essays Res · Sep 2014

    A comparative study between Truview(PCD) laryngoscope and Macintosh laryngoscope in viewing glottic opening and ease of intubation: A crossover study.

    • Sourav Kr Bag, V R Hemanth Kumar, N Krishnaveni, M Ravishankar, J Velraj, and M Aruloli.
    • Department of Anaesthesiology and Critical Care, Mahatma Gandhi Medical College and Research Institute, Pillaiyarkuppam, Puducherry, India.
    • Anesth Essays Res. 2014 Sep 1;8(3):372-6.

    BackgroundTruview(PCD) laryngoscope is specially designed to aid positioning of the endotracheal tube as well as to record entry of the tube into glottis. Aim of the study is to compare the view of glottic opening and ease of intubation between Truview(PCD) laryngoscope and Macintosh laryngoscope in patients undergoing general anesthesia (GA).MethodologyTwo hundred patients undergoing elective surgery, under GA were randomized into two groups, Group TV and Group ML. In Group TV, Trueview(PCD) laryngoscope was used initially to visualize the vocal cords for Cormack and Lehane grading (CLG) and to spray the vocal cords with 10% lignocaine. Then the patient was ventilated for 1 min and Macintosh laryngoscope was used to visualize the vocal cords for CLG and proceed with intubation. In Group ML, Macintosh laryngoscope was used initially and later Truview(PCD) laryngoscope. Time to intubation, CLG, number of attempts and hemodynamic parameters were recorded.ResultsNinety-six and 89 patients had CLG1 visualization when Truview(PCD) laryngoscope was used as 1(st) and 2(nd) device respectively compared to 41 and 68 with Macintosh laryngoscope (P = 0.00). Four patients had CLG 4 visualization with Macintosh laryngoscope that turned out to be grade II visualization with Truview(PCD) laryngoscope (P = 0.00). Mean time taken for intubation with Truview(PCD) and Macintosh laryngoscope was 21.10 ± 5.64 s and 15.79 ± 2.76 s respectively (P = 0.00).ConclusionBetter visualization with lesser CLG was found with Truview(PCD) laryngoscope but it took longer time for intubation than Macintosh laryngoscope. The hemodynamic response to intubation was significantly less with the use of Truview(PCD) laryngoscope when compared to that of Macintosh laryngoscope.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…