Anestezjol Intens Ter · Apr 2008
Comparative Study[Carbon monoxide pulse oximetry vs direct spectrophotometry for early detection of CO poisoning].
- Jacek Kot, Zdzisław Sićko, and Piotr Góralczyk.
- Klinika Medycyny Hiperbarycznej i Ratownictwa Morskiego, Akademickie Centrum Medycyny Morskiej i Tropikalnej, AM w Gdańsku. jkot@amg.gda.pl
- Anestezjol Intens Ter. 2008 Apr 1;40(2):75-8.
BackgroundDirect spectrophotometry has been recognized as a standard reference method in the treatment of victims of carbon monoxide (CO) poisoning. Recently, Masimo (USA) has developed a pulse oximeter for the detection of HbCO. Using new probes, similar to those used for traditional pulse oximetry, several different light wavelengths enable detection of different haemoglobins.MethodsWe have compared forty-nine capillary blood samples taken from patients admitted to the hyperbaric center with CO poisoning. The samples were analyzed using direct spectrophotometry (HbCO) and compared with the corresponding pulse CO-oximeter (SpCO) readings. The Bland-Altman method was used for statistical analysis.ResultsThe mean HbCO concentration was 18.1+/-12.7% (range 0.1% to 47.4%) and the mean SpCO concentration, 17.6+/-11.3% (range 1.0% to 46.0%). There was a strong positive correlation between laboratory results and bedside readings (r2=0.88). The mean difference between readings was 0.5+/-4.3% (range -11.0% to +9.0%), and the distribution was uniform over the whole range of measured levels. For detection of HbCO levels higher than 20%, the sensitivity of the pulse-CO-oximeter was 77.8%, PPV 82.4%, specificity 90.3%, and NPV 87.5%.ConclusionsOur study confirmed the accuracy of pulse CO-oximetry for rapid detection of the presence and concentration of HbCO. Since this method is based on analysis of peripheral blood flow, it is not clear if it could be used in patients with low perfusion, hypothermia or burns.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.