Performance of a reservoir nasal cannula (Oxymizer) during sleep in hypoxemic patients with COPD.
- E M Hagarty, M S Skorodin, W M Stiers, M B Mamdani, J A Jessen, and E C Belingon.
- Edward Hines Jr. Hospital, Department of Veterans Affairs, Hines, IL.
- Chest. 1993 Apr 1;103(4):1129-34.
Study ObjectiveTo determine whether a reservoir nasal cannula (RNC) (Oxymizer) provides an arterial hemoglobin oxygen saturation as measured by pulse oximetry (SpO2) equivalent to that provided by the standard nasal cannula (SNC) during sleep in hypoxemic patients with COPD while reducing oxygen flow requirement and cost.DesignThe study took place in a sleep laboratory for three nights, with the first night for acclimatization to the new sleeping environment. In a repeated-measures design, on the second and third nights, subjects used the SNC for one night and the RNC on another night. The order in which they received the two devices was counterbalanced.SubjectsThe subjects were patients with COPD who had a stable PaO2 of 55 mm Hg or less or had a value of 56 to 59 mm Hg with evidence of cor pulmonale or polycythemia (or both) and an FEV1/FVC of less than 70 percent.InterventionsA pulse oximeter was used to measure SpO2. An arterial blood gas measurement was taken on each night while the patients with COPD were receiving oxygen therapy via the assigned device. An EEG machine was used to record measurements of electro-oculography, chin electromyography (EMG), anterior tibialis EMG and EEG.Measurements And Main ResultsThere was a statistically significant difference between mean SpO2 during sleep (RNC, 91 percent; SNC, 93 percent; F = 7.89; p = 0.01). Nocturnal SpO2 was less than 90 percent for 24.2 percent of the time with the RNC and for 17.5 percent of the time with the SNC (F = 5.41; p = 0.03), but there was no significant difference in the amount of time that SpO2 was less than 85 percent. Compared to the SNC, in 4 of 26 patients with COPD, the RNC performed better; in 12 patients with COPD, the RNC performed the same, and in 10 patients with COPD the RNC performed worse during sleep. Sleep parameters were not significantly different between the two devices.ConclusionsThe difference of 2 percent in mean SpO2 is within the range of SpO2 measurement error. Therefore, the two devices are equally effective when the sample is considered as a whole. Nighttime oximetry is necessary prior to prescription, since nighttime efficacy of the RNC cannot be predicted on the basis of daytime pulse oximetry.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.