    Variability in infusion pressure and continuous flow rate delivered from pressurized bag pump flush systems.

    • M I Hug, V Buettiker, A Cornelius, and M Weiss.
    • Department of Anaesthesia, University Children's Hospital Zurich, Switzerland.
    • Anaesth Intensive Care. 2002 Jun 1;30(3):341-7.

    AbstractUsing 10 different infusion bag pressure pumps, indicated manometer pressures were compared with measured infusion pressures proximal to the flow-regulating device in an in vitro experiment. Flow rates delivered through the flow-regulating device were gravimetrically measured at different monitored pressure levels. Significant differences were found between manometer and measured infusion pressures among the tested pressure bagpumps (e.g. 500 ml bag volumepressurized to 300 mmHg manometerpressure: 219.6 +/- 7.8 to 407.2 +/- 2.7 mmHg). The infusion pressures were additionally affected by the vertical level of the infusion bag pump and by the volume of the infusion bag. Flow rates delivered through the flow-regulating device were directly correlated to the measured infusion pressure (r2= 0.9926). Differences inflow rates can have a considerable impact on maintaining catheter patency and avoidance of fluid overload and retrograde flushing into the central arterial circulation in neonates and small children. A simple manoeuvre using the invasive pressure transducer allows monitoring and adjustment of the infusion pressure in the clinical setting.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…