    Accuracy of three transcutaneous carbon dioxide monitors in critically ill children.

    • Javier Urbano, Verónica Cruzado, Jesús López-Herce, Jimena del Castillo, José María Bellón, and Angel Carrillo.
    • Research Assistant Contract, Spanish Health Research Fund (Fondo de Investigaciones Sanitarias), Instituto de Salud Carlos III, Madrid, Spain. javierurbano@mixmail.com
    • Pediatr. Pulmonol. 2010 May 1;45(5):481-6.

    ObjectivesTo study the accuracy of three devices for measuring transcutaneous CO(2) tension in critically ill children.MethodsA prospective study comparing the values from three transcutaneous CO(2) monitors (SenTec, TOSCA 500, and TINA TCM3) with simultaneous arterial CO(2) (PaCO(2)) and end-tidal CO(2) (EtCO(2)) values. Clinical data were collected from the patients. Influence of core-skin temperature gradient and doses of inotropic drugs was evaluated.ResultsThere were 62 samples from 41 critically ill children with ages between 2 and 192 months (median, 18.5 months) and weights between 3.1 and 72 kg (median, 9 kg). The median PaCO(2) was 42.5 mmHg (range, 28-85 mmHg). Transcutaneous CO(2) (PtcCO(2)) values correlated better with PaCO(2) than with EtCO(2). The correlation coefficient between PaCO(2) and PtCO(2) was 0.833 with the TINA TCM3 monitor, 0.931 with the SenTec monitor, and 0.765 with the TOSCA 500 monitor. The mean (SD) differences between the PaCO(2) and PtcCO(2) were 4.5 (3.7) mmHg, 4.3 (3.8) mmHg, and 5.6 (5.1) mmHg, respectively, with the three monitors, and the differences between the PaCO(2) and PtcCO(2) were less than 7.5 mmHg in 77.7%, 81.2%, and 67.7% of the samples. Bland-Altman analysis showed a precision of +/-11.5 mmHg for TINA TCM3 monitor, +/-10.6 mmHg for SenTec monitor, and +/-14.8 mmHg for TOSCA monitor. No influence of core-skin temperature gradient and inotropic index on the differences between PaCO(2) and PtcCO(2) was observed.ConclusionsThe three transcutaneous CO(2) monitors have an acceptable correlation with arterial CO(2) tension and can be useful in critically ill children.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.